3. Check CCD cleaning

Finding: It turns out we have no control over CCD "cleanings" in this operational mode.

These graphs should check CCD cleaning. See excerpt from Tasks doc below.

  • Resonon tasks_cjcomments.doc

    Tests 1 and 2 below should be done each time the setup is moved or significantly reconfigured, though in the future only one or two long integration time FDI scans would likely be required to confirm the absence of light leakage.


    File name Experiment
    check CCD cleaning(He/Hg light source on, all fibers illuminated)
    3c_10BI_01_ Take 10 BI images @ 0.35 second integration time (Shutter Control Disabled Closed)
    3c_10LI_ Take 10 LI images @ 0.35 second integration time (Shutter Control Normal)
    3c_10BI_02_ Take 10 BI images @ 0.35 second integration time (Shutter Control Disabled Closed)
    3d_cleaning_1_ Collect 1 image set at CCD cleaning settings 1
    3d_cleaning_5_ Collect 1 image set at CCD cleaning settings 5
    3d_cleaning_10_1 Collect 1 image set at CCD cleaning settings 10


    Subject: RE: CCD function
    Date: Fri, 11 Dec 2009 13:59:08 -0500
    From: Lichty, Alan
    To: Mark Yarbrough


    Hi,
    from talking to engineering, it is not a design flaw but a design choice. There are pros and cons for others ways to clean and we try to go the middle of the road because other applications suffer is we do the cleans differently.

    If you would like to have a conference call with our product manager let me know. He can explain what we can and can’t do with cleans and the reasoning behind it.

    Sincerely,

    Alan Lichty





    pwd: C:\zflora\mldata\MOBY2\Intital_testing\091208
    Date: 14-Dec-2009 12:06:41
    Created from marktasks_3(1)