Hi Steph,
Riley & I managed to collect some BS04cfg101 & RS03cfg001 data yesterday,
which are at /ftp1/Mike/HI-2019-03/ = ready for your data processing help.
I wanted to setup both specs to collect data somewhat-simultaneously -
I'm not sure that doing one-spec-at-a-time was such a bad idea now...
There are photos of the setup, in the cal van next to the tent.
Both specs' bare fiber-bundle fibers were looking into the Spectralon sphere.
There were no internal aux data (degC, %RH) collected from the specs.
Day01 had as input to the sphere a Thorlabs LED. I think this was the first use
of this source. Carol found this in 2015 and we bought one = LED driver /
power supply, aka Thorlabs DC2100, plus 5x different LED "heads". Last night
we scanned the M530L3 head, which should be centered ~ 530 nm.
The LED pwr sup was set at 9 mA ( == LOW power ), as noted in the log sheet.
I ran the Labsphere Si PhotoDiode monitor at the sphere monitor port #A.
This would be the PD that Carol had at NIST for cals (which you should have
data from?) but, the detector was not a UV-enhanced kine, so it has no
response in the UV, and we will try to re-purchase another PD with a
UV-enhanced detector. The PD's electrical-current to voltage pre-amp was
the one Carol had at NIST for the NIST cals, aka Femto DLPCA-200. The Voltage
output from the pre-amp was logged via the 34970A DVM using software
Vdc6_5.vi, which you have seen output from before. Someday I will update
this VI for use with a PD monitor... Last night, the PD preamp went to DVM
chan#3 (i.e. setup for high precision Voltage measurement, typically for a
shunt during Ed/Lu absolute lamp cals), and DVM chan#5 had the Radio Shack
10 kOhm thermistor for ambient temperature monitor. For now the only
indication of what gain setting the PD preamp was set at is noted in the
log sheet - last night was @ low gain, 10^4. This will someday be included
in the log file's header...
The log file from the DVM is in the HI-2019-03/BSRS/ subdir on the NAS.
The 1x day01 log sheet is in the HI-2019-03/doc/ subdir on the NAS.
One problem last night: I did not power ON the TEC for the PD monitor.
Tonight I will try harder to turn that bugger on.
We had to re-load the Andor SOLIS software onto 2x HP laptops to control the
BS & RS. So I'm hoping that I set all acquisition parameters correctly. The data
acq sequence was 2x bac, 4x sig, 2x bac.
Tonight we will attempt to collect some Ekspla laser scans via BS & RS with
all bare bundle fibers illuminated.
Good Luck with this! MF
|