MOBYrefresh - Configurations


MOBY-Refresh will have 4 each blue and red specs - BS01, BS02, BS03, BS04 and RS01, RS02, RS03, RS04
Page#
SPEC LinkDescription

Date Added

1.01 BS01 timeline PDF MOBYrefresh BS01 information document 4/27/2015
1.02 All Resonon filtering, v4 Status of optical filters in the Resonon BS & RS spectroradiometers  
         
         
         

 

Resonon Spectrograph configuration pages - shows Resonon data and experiments run on each spec
SPEC Link Description

Date Added

BS01 config page MOBYrefresh BS01 configuration pag 4/20/2016
BS02 config page MOBYrefresh BS02 configuration page 4/20/2016
BS03 config page MOBYrefresh BS03 configuration page 2/07/2016
BS04 config page MOBYrefresh BS04 configuration page 4/29/2016
RS01 config page MOBYrefresh RS01 configuration page 4/20/2016
RS02 config page MOBYrefresh RS02 configuration page 4/20/2016
RS03 config page MOBYrefresh RS03 configuration page 8/09/2016
RS04 config page MOBYrefresh RS04 configuration page 9/12/2016
MarONet has the BS and RS 05 an 06, to see those config files here

 

Resonon Spectrograph configuration summary
SPEC Pre-Delivery Doc Wavelength Range
min- max

Andor iKon-M Camera
SN

Smile
@ 388 and 587 nm

(pix less than)

Keystone (Peak to Trough)
Channel 1/7/14

(less than)

Spectral Resolution
(FWHM)
@ 388 and 587 nm

Grating

Batch

Stability
(“shake” test)

BS/RS

together

UV ghost &

wedge window

Other
BS01cfg100 Yes 340- 697 17878 < 1 & 0.2 1.5 / 1.5 / 2 < 1.6 & < 0.9 Aug. 2016 1 pix No No Shake showed too much movement so they tighted screws and redid the shake test
BS01cfg101 Yes 339.78-696 17878 < 1 & 0.2 1.5 / 1.5 / 2 < 1.6 & < 0.9 Aug. 2016 0.4 pix No No After second shake test the result was much better
BS04cfg100 Yes 347.58-705.16 17881 1 pix 1/ 1/ 2 pixs 1.2 & 1.6 Aug. 2016 - No Fixed Returned to Hawaii after wedge window fix at Andor and re-assembly at Resonon - 19 Jun 2019. This config includes Resonon data
BS04cfg101                   UV ghost found??? Hawaii-2019-03: testing how assembly, found a UV ghost on some data
RS01cfg001 Yes 504.46-915.69 17879 0.3 & 0.3 1.5 / 1.5 / 4.0 ~ 1 & ~1 ? - No - Sent to Andor for window repair, then Resonon and they arrived in HI in good shape on 23-Aug-2019
RS02cfg001 Yes 489.09-901.0 17880 0.3 & 0.2 1 / 1.5 / 4.5 pixels ~ 1.1 nm Aug. 2015 < 0.1 nm No Yes

17 Oct 2019: Arrived in Hawaii in visually in good condition

 

Camera documents - current configs
Andor iKon-M Camera Link

Mike Excel sheet

White/Black spots (#) Uniformity/linearity/rsp linearity Min Dark current acheivable Saturation Signal per pixel Window Resonon Spec Camera
Model
Other

Date Added

16343* PDF xlsx ??RS05 1 0.2816 / 1 / 2.2 0.000508 127747 Broadband_VUV_NIR_Wedged RS05 DU934P-BR-DD   2/14/2019
16344* PDF   1 0.285 / 1 / 1.74 0.000568 133552 Broadband_VUV_NIR_Wedged   DU934P-BR-DD   2/14/2019
16345 PDF   1 0.5760 / 1.00 / 2.22 0.00089200 119380 AR_coated_Window(1/2 deg wedge)   DU934P-BR-DD   2/14/2019
16346 PDF   0 0.2742 / 1.00 / 1.75 0.00069000 120022 AR_coated_Window (1/2 deg wedge)   DU934P-BR-DD   2/14/2019
17878* PDF   6 0.401 / 1 / 1.90 0.000581 106013 Broadband_VUV_NIR_Wedged   DU934P-BR-DD   2/14/2019
17879* PDF   1 0.29705 / 1 / 1.85 0.000732 134543 Broadband_VUV_NIR_Wedged   DU934P-BR-DD   2/14/2019
17880* PDF   0 0.2383 / 1 / 1.86 0.000578 147857 Broadband_VUV_NIR_Wedged   DU934P-BR-DD   2/14/2019
17881* PDF xlsx 0 0.238 / 1 / 1.89 0.000582 138015 Broadband_VUV_NIR_Wedged BS04cfg100 DU934P-BR-DD   2/14/2019
19347 PDF   8 0.3944 / 1.00 / 1.90 0.00068700 128246 Broadband_VUV_NIR_Wedged   DU934P-BR-DD   2/14/2019
19348 PDF   0 0.6060 / 1.00 / 1.89 0.00078000 126243 Broadband_VUV_NIR_Wedged RS03cfg002 DU934P-BR-DD   2/14/2019
22932* PDF   0 0.2978/ 1 / 2.22 0.00099 151072 Broadband VUV-NIR Wedged   DU934P-BR-DD

 

2/14/2019
20477* PDF   0 0.3788 / 1 / 1.70 0.000522 140999 Broadband VUV-NIR Wedged  

DU934P-BR-DD-9XT

  2/14/2019

To see the cameras that where removed and returned to Andor for repair and addition of a wedged window see table below
* Camera parallel window to be replaced with WN35FS Broadband VUV-NIR Wedged windows, code- (88-W-NR)W. In May 2019 all cameras with fixed windows where sent to Resonon to repair then with their spec.
Mikes Excel Docs (Resonon_re-assembly_notes.pdf & Resonon_re-assembly_notes.xlsx) (FedEx shipments to Resonon, to re-assemble the specs w/ Andor cameras after the Andor repairs for wedged windows.).
The BS01 spec (camera 17878) had shifting track positions on the CCD during deployment in MOBY262 (Feb to Aug-2017), so Ken was asking if this spec could be opened to look for a loose element.

Andor Camera Software Document

Components use in MOBrefresh


Resonon Spectrograph configuration summary
SPEC Pre-Delivery Doc Wavelength Range
min- max

Andor iKon-M Camera
SN

Smile
@ 388 and 587 nm

(pix less than)

Keystone (Peak to Trough)
Channel 1/7/14

(less than)

Spectral Resolution
(FWHM)
@ 388 and 587 nm

Grating

Batch

Stability
(“shake” test)

BS/RS

together

UV ghost &

wedge window

Other
Below are the old configurations after some specs had the wedge window repair we are starting all new cfgs at 100
BS01cfg001 No 346.13 - 704.29 16346 - - - Dec. 2014 - No No 14-Mar-2015 arrived in HI. Iinfrared artifact was found. Hawaii-2015-01
BS01cfg002 Yes 339.88- 696.68 17878 0.6 and 0.5 pixs 1 / 1 / 2 pixs < 1.0 and 1.1 nm Dec. 2014 <0.1 nm No Yes Apr-2015 Casey Smith modified slit mount to decrease the distance between the fiber face and the slit. Hawaii-2015-03
May-2015
Mike Kehoe found the source of the "infrared Artifact" (and wrote Report...docx). New camera.
BS01cfg003   340.15 - 697.61   - - -   - No Yes Feb-2016 Resonon added new filter at tilted substrate (replaces AR coated silica) to remove Infrared Artifact, filter = Chroma ET695sp 8deg lot#300205.
Hawaii-2016-02
Took data before shipping to Miami.
BS01cfg004       - - -   - No Yes Apr-2016: Arrived in Miami with loose resonon tube shipping, Ken decided there was no damage. Miami-2016-01
BS01cfg005       - - -   - No Yes Sent from Miami Back to Resonon. Resonon replaced the the Schrader valve on BS01, and realigned the unit. Mike replaced the screws that hold the camera mount in place with slightly longer ones, and when I was tightening them as one of them felt a bit 'mushy'. No data where collected at Resonon. Then shipped to Mike in Hawaii. 14-Sep-2016 Arrived in Hawaii: Mike says "The condition of the packing job was not optimal." Hawaii-2016-09.
Feb 2017 -
preped to go in MOBY262
BS01cfg006       - - -   - No Yes Hawaii-2017-01: day 1 adding iris and reducing light signal.
UV ghosting from mirrored tracks - Hawaii-2017-01 - Page number 5.06
BS01cfg007       - - -   - No Yes Hawaii-2017-01 day 2-4 and MOBY262: Final data set on iris adjustments and in-water data, MOBY-L264 postcals
BS01cfg008       - - -   - No Yes MOBY-L267: checking for UV ghost post deploymnet
BS02cfg001 Yes 342.85 - 701.46 16345 0.2 and 0.5 pixs 1 / 1 pixs < 0.7 and 0.8 nm Dec. 2014 no change No No 02-Feb-2016 arrived in Hawaii. Hawaii-2016-01
BS02cfg002 -               No No 09-May-2016: shutter block added, controller installed, new LabJack aux data aqc running.
Hawaii-2016-01: NO UV ghosting from mirrored tracks - Page number 3.03
BS02cfg003 -               No No Jul-2016: BS02cfg003 was installed in the MOBY261 buoy and deployed on 16 Aug. BS02 was connected to duplicate Lu, Ed and Es collectors on the buoy for in-water data collection. 03-April-2017: the BS02 was removed from the M261 housing so it's camera CCD-16345 could be sent back to Andor for repairs.
BS03cfg001 Yes 337.80 - 697.04 16346 < 1 2/ 2 / 2 pixs ~1.3 and < 1 Aug. 2016 <0.1 nm No No 09-Feb-2017 arrive in Hawaii. Data collection during Hawaii-2017-02
BS03cfg002               No No Hawaii-2017-05 pre testing (July 2017) for MOBY263 deployment (Aug 2017) - day01. NO UV ghosting from mirrored tracks - Page number 4.03
BS03cfg003               No No

Hawaii-2017-05 pre testing (July 2017) for MOBY263 deployment (Aug 2017) - day02. NO UV ghosting from mirrored tracks - Page number 4.03
MOBY263
deployment. MOBY-L267 postcals.

BS03cfg004               No No

???

BS03cfg005               Yes No

MOBY265 deployment (BS/RS04). MOBY-L274 postcal.

BS03cfg006               Yes No

Hawaii-2019-01 (BS/RS04)

BS03cfg007               Yes No

MOBY-L279 - pre_MOBY267 cal and assembly - bare fiber

BS03cfg008               Yes No

MOBY-L279 - pre_MOBY267 cal and assembly - 4x FO attached, FO block, Es no filters

BS03cfg009               Yes No MOBY-L279 - pre_MOBY267 cal and assembly - ND filteers added
MOBY267 - installed with BS4, flooding of LPJ killed the BS/RS system no data collected
BS04cfg001 Yes 340.26 - 698.14 17881 1 pixs 1/ 1/ 2 pixs < 1 nm Dec. 2014 <0.1 nm No Yes Filter on front face of front prism, Filter on rear tilted substrate.
18 Jul 2016: Arrived in Hawaii.
Hawaii-2016-08: first data set
MOBY267: bare fiber
BS04cfg002                 No Yes MOBY-L267 precals, Dec 14 = shutters installed and iris all tracks, precals
BS04cfg003                 No Yes MOBY-L267: Dec 15 = Mark Iris removed from Track 1 and Mike tightened down the locking screws for the 4x irises I adjusted yesterday
BS04cfg004                 No Yes MOBY-L267, Dec 16 = added the "window plate" to the BS shutter block, then coupled the window plate to the FO block in the tent. UV ghosting from mirrored tracks - Page number 9.04
BS04cfg005                 No Yes MOBY264: Final assembly in MOBY after system response done then on 19 Dec 2017: deployed on MOBY264.
MOBY-L271: postcals
Camera was remvoed and send bakc to Andor for window repair and then Resonon for refocusing. Next cfg# will start at 100
SPEC Pre-Delivery Doc

Wavelength Range
min- max

Andor iKon-M Camera

Smile
@ 587 and 705 nm

Keystone (Peak to Trough)
Channel 1/7/14

Spectral Resolution
(FWHM)
@ 587 and 705 nm

Grating

Batch

Stability
(“shake” test)

BS/RS

together

UV ghost &

wedge window

Other
RS01cfg001 Yes 503.83 - 911.79 17879 0.25 and 0.50 pixs 0.5/ 0.5 / 5.0 pixels < 0.7 and 0.6 nm Aug. 2015 < 0.1 nm No Yes Apr-11-2016 Arrived in HI. 540-900 nm bandpass filter
Hawaii-2016-03. Hawaii-2016-07 - testing of the software pre-MOBY262 .
MOBY-L267, UV ghosting from mirrored tracks - Page number 2.04
RS02cfg001 Yes 487.15 - 895.75 17880 0.7 and 0.6 pixs 3 / 0 / 5 pixels < 1 nm Aug. 2015 < 0.1 nm No Yes

18 Jul 2016: Arrived in Hawaii
540-900 nm bandpass filter. Shipped to Miami on 12 Feb 2018.
Sent bay to Hawaii by Art (dates??).
MOBY-L267
, Initial data, UV ghosting from mirrored tracks - Page number 2.03

RS03cfg001 Yes 499.46 - 907.71 19347 0.4 and 0.3 pixs 1.5 / 1.0 / 5.0 pixels < 1 nm Aug. 2015 < 0.1 nm No No

8 Aug 2016: Arrived in Hawaii
540-900 nm bandpass filter: Hawaii_2016-05, initial testing,
NO UV ghosting from mirrored tracks - no data but assumed because this spec as a wedged window

RS03cfg002                   UV ghost found??? Shipped to Miami on 25 Aug 2016. Miami-2016-02 - software test
13 May 2019: arrived in Hawaii in fein shape.
Hawaii-2019-03: testing how assembly, found a UV ghost on some data
RS04cfg001 Yes 499.24 - 907.19 19348 0.6 and 0.3 pixs 0.5 / 1.5 / 6 pixels < 1 nm Aug. 2015 < 0.1 nm No No

Arrived in Hawaii 14 Sep 2016
MOBY-L271,inital data
NO UV ghosting from mirrored tracks - no data but assumed because this spec as a wedged window

RS04cfg002                 Yes No

MOBY265, MOBY-L274 postcals (BS03/RS)

RS04cfg003                 Yes No

Hawaii-2019-01, fiber bending (BS03/RS)

RS04cfg004                 Yes No

MOBY-L279 - pre_MOBY267 cal and assembly - bare fiber

RS04cfg005                 Yes No

MOBY-L279 - pre_MOBY267 cal and assembly - 4x FO attached, FO block, Es no filters

RS04cfg006                 Yes No MOBY-L279 - pre_MOBY267 cal and assembly - ND filteers added
MOBY267 - installed with BS4, flooding of LPJ killed the BS/RS system no data collected

 

Camera documents - OLD CAMERA CONIGS before being sent back to Andor to have the wedged windows installed (There cameras where removed and returned to Andor for repair. To place the windows with a wedged window (~2018 July-Aug))
By Feb 2019 all the Camera where back from Andor and In Hawaii again.

Andor iKon-M Camera Link White/Black spots (#) Uniformity/linearity/rsp linearity Min Dark current acheivable Saturation Signal per pixel Window Resonon Spec Camera
Model
Other

Date Added

16343 PDF 0 0.4380 / 1.00 / 2.18 0.00071700 129050 none specified MOBYnet DU934P-BR-DD   4/21/2016
16344 PDF 1 0.4380 / 1.00 / 2.18 0.00071700 129050 none specified MOBYnet DU934P-BR-DD   4/21/2016
17878 PDF 6 0.4100 / 1.00 / 1.89 0.00090000 107790 Broadband_VIS_NIR_Parallel BS01cfg002,3 and 4 DU934P-BR-DD   4/20/2016
17879 PDF 1 0.4163 / 1.00 / 1.85 0.00056200 137685 Broadband_VIS_NIR_Parallel RS01cfg001 DU934P-BR-DD   4/20/2016
17880 PDF 0 0.3763 / 1.00 / 1.87 0.00056200 140609 Broadband_VIS_NIR_Parallel RS02cfg001 DU934P-BR-DD   4/21/2016
17881 PDF 0 0.2589 / 1.00 / 1.89 0.00065600 70237 Broadband_VIS_NIR_Parallel BS04cfg001 DU934P-BR-DD   4/21/2016
20476 PDF 0 0.4298 / 1.00 / 1.7 0.000423 129259 Broadband VIS-NIR Parallel MOBYnet DU934P-BR-DD-9XT Arrival Photos and notes, HAW16-03 data 3/24/2017
20477 PDF 0 0.4125 / 1.00 / 1.75 0.000489 146896 Broadband VIS-NIR Parallel  

DU934P-BR-DD-9XT

Arrival photos and notes 4/20/2017