MOBYnet - Stability - Documents
MOBY-net will have 2 each blue and red specs - BS5 and BS6 and RS5 and RS6
A radiometer (CAS, Instrument Systems) and a stability source (SQM-5002, Yankee Environmental Systems) have been purchased for the MOBY-NET project. The idea is to ship the monitor system with the optical buoy between deployments (somewhere besides Hawaii) and calibration (at the MOBY Hawaii site). We will be doing the characterization here, and I will also do some calibrations so we can estimate the levels as well as track what may be changing with independent sources or radiometers.
Doc#
| Doc Name and Link | Person | Description | Date Updated | Spec |
|---|---|---|---|---|---|
| 200.01 | SQM Manual (PDF) | Carol | SQM-5002 Satellite Quality Monitor Installation and User Guide | 26 Mar 2015 | SQM |
| 200.02 | CAS brochure (PDF) | Carol | CAS 140CT Array Spectrometer brochure | 26 Mar 2015 | CAS |
| 200.03 | Test Plan (docx) | Carol | Draft test plan for MOBY NET Stability Testing at NIST |
26 Mar 2015 | plan |
| 200.04 | Update report (PPTx) | Carol | SQM & CAS Status Report | 4 Aug 2015 | update |
| 200.05 | MODEL SQM-5002 | Carol | SATELLITE QUALITY MONITOR MODEL SQM-5002: Overboew | 18 Aug 2015 | SQM |
| 200.06 | SIMBIOS paper | Carol | Long Term Monitoring of Radiometer Sensitivity for Radiometric Comparisons among Optical Laboratories | 18 Aug 2015 | SQM |
| 200.07 | SQM paper | Carol | Calibration Evaluation and Radiometric Testing of Field Radiometers with the SeaWiFS Quality Monitor | 18 Aug 2015 | SQM |
| 200.08 | SQM paper | Carol | The SeaWiFS Quality Monitor· a portable field calibration light source | 18 Aug 2015 | SQM |
| 200.09 | SQM paper | Carol | Simulation of a Compact Large-Area Radiometric Light Source | 18 Aug 2015 | SQM |
| 200.10 | SQM paper | Carol | Radiometric and Engineering Performance of the SeaWiFS Quality Monitor (SQM): A Portable Light Source for Field Radiometers | 18 Aug 2015 | SQM |
| Problems Seen | |||||
| 300.01 | CAS problem 1 | Carol | Foreoptic selections for the MOBY NET CAS. Solution: we have ordered two fiber bundle systems – irradiance and bare fiber; TBD design radiance head | 3 Sep 2015 | CAS |
| 300.02 | SQM Problem 1 | Carol | SQM housekeeping logging software. Solution: we have put YES’ V0 into Labview 2013 and call this V1. After deciding operating protocols, we will make a V2 | 3 Sep 2015 | SQM |
| 300.03 | SQM Problem 2 | Carol | SQM adaptor plate for holding the radiometers to be tested. Solution: we will keep the YES plate and use its 4.125” ID hole to mate to. | 3 Sep 2015 | SQM |
| 300.04 | SQM Problem 3 | Carol | New SQM Manual and Notes as we go along. Solution: we making notes in the manual or the web site. | 3 Sep 2015 | SQM |
| 300.05 | SQM Problem 4 | Carol | :No spare SQM lamps. Solution: we ordered replacement lamps and will make the mounting fixtures and pot them in the NIST shops. | 9 Sep 2015 | SQM |
| 300.05 | SQM Problem 5 | Carol | There is little UV because the window is glass and the diffuser is acrylic. Solution: replace both with fused silica (quartz)-based materials. | 17 Sep 2015 | SQM |
| 300.06 | CAS thermal test doc | Carol | Temperature Profiles during CAS Thermal Tests | 14 Feb 2019 | CAS |